DOI 10.1055/s-0041-1735551

Медицинская библиотека · Veterinary and comparative orthopaedics and traumatology : V.C.O.T · 2021

Influence of Screw-Hole Defect Size on the Biomechanical Properties of Feline Femora in an Ex Vivo Model.

Hoon QiCai Jason, Wang Tian, Hall Evelyn, Walsh William R, Johnson Kenneth A

СтатьяRandomized Controlled Trial, VeterinaryПеревод в очереди

Аннотация

OBJECTIVE

The study aims to evaluate the biomechanical properties of feline femora with craniocaudal screw-hole defects of increasing diameter, subjected to three-point bending and torsion to failure at two different loading rates.

STUDY

DESIGN

Eighty femoral pairs were harvested from adult cat cadavers. For each bending and torsional experiment, there were five groups (n = 8 pairs) of increasing craniocaudal screw-hole defects (intact, 1.5 mm, 2.0 mm, 2.4 mm, 2.7mm). Mid-diaphyseal bicortical defects were created with an appropriate pilot drill-hole and tapped accordingly. Left and right femora of each pair were randomly assigned to a destructive loading protocol at low (10 mm/min; 0.5 degrees/s) or high rates (3,000 mm/min; 90 degrees/s) respectively. Stiffness, load/torque-to-failure, energy-to-failure and fracture morphology were recorded.

RESULTS

Defect size to bone diameter ratio was significantly different between defect groups within bending and torsional experiments respectively (intact [0%; 0%], 1.5 mm [17.8%; 17.1%], 2.0 mm [22.8%; 23.5%], 2.4 mm [27.8%; 27.6%], 2.7 mm [31.1%; 32.4%]) (p < 0.001). No significant differences in stiffness and load/torque-to-failure were noted with increasing deficit sizes in all loading conditions. Screw-hole (2.7 mm) defects up to 33% bone diameter had a maximum of 20% reduction in bending and torsional strength compared with intact bone at both loading rates. Stiffness and load/torque-to-failure in both bending and torsion were increased in bones subjected to higher loading rates (p < 0.001).

CONCLUSION

Screw-hole defects up to 2.7 mm did not significantly reduce feline bone failure properties in this ex vivo femoral study. These findings support current screw-size selection guidelines of up to 33% bone diameter as appropriate for use in feline fracture osteosynthesis.

Для цитирования: Hoon QiCai Jason, Wang Tian, Hall Evelyn, Walsh William R, Johnson Kenneth A. Influence of Screw-Hole Defect Size on the Biomechanical Properties of Feline Femora in an Ex Vivo Model.. Veterinary and comparative orthopaedics and traumatology : V.C.O.T. 2021. DOI: 10.1055/s-0041-1735551.