Length of Plates and Number of Screws for the Fixation of Distal Humerus Fractures: A Finite Element Biomechanical Study.

Zarifian Ahmadreza, Akbarinezhad Fough Ali, Eygendaal Denise, Rivlin Michael, Mousavi Shaegh Seyed Ali, Kachooei Amir R

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: We performed a biomechanical analysis using the finite element method to assess the effects of plate length and the number of screws on construct stiffness, stress distribution, and fracture displacement in the fixation of type A2 distal humerus fractures.

METHODS

A 3-dimensional humerus model was constructed using computed tomography of a healthy man. After creating a 2-mm extra-articular fracture gap, orthogonal double-plate fixation was performed with an incremental increase in plate length and the number of screws, creating 17 fixation models. Four screws were placed in each plate's distal segment, and the number of screws was increased incrementally in the segment proximal to the fracture, starting from 2 in the medial (M) and 2 in the lateral (L) plate (M2∗L2).

RESULTS

The fifth screw proximal to the fracture in the lateral plate (L5) played an essential role in increasing stiffness under bending, axial, and torsional forces surpassing the intact bone, which may have been due to the bypassing of the stress riser area. Minimum construct stiffness was created when 5 (M3∗L2) screws were inserted into the proximal segment. For bending forces, the M4∗L2 construct was stronger than M3∗L3 (total 6 proximal screws), and M5∗L3 was stronger than M4∗L4 (total 8 proximal screws), showing higher stiffness when the plates ended at different levels. The M4∗L2 construct (6 screws) had stiffness comparable with M4∗L3, M4∗L4, and M5∗L4 during bending, showing comparable stiffness with the least instrumentation density.

CONCLUSION

Our findings suggested M3∗L5 as the optimum and M3∗L2 as the minimum construct to resist all bending, axial and torsional forces.

CLINICAL RELEVANCE: Applying the results may improve surgical techniques, decrease the rate of complications, including fixation failure and nerve injury, and optimize the time of surgery. Moreover, hardware removal is less cumbersome with fewer screws.

Для цитирования: Zarifian Ahmadreza, Akbarinezhad Fough Ali, Eygendaal Denise, Rivlin Michael, Mousavi Shaegh Seyed Ali, Kachooei Amir R. Length of Plates and Number of Screws for the Fixation of Distal Humerus Fractures: A Finite Element Biomechanical Study.. The Journal of hand surgery. 2021. DOI: 10.1016/j.jhsa.2021.07.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.