MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.injury.2009.08.014

Медицинская библиотека · Injury · 2009

Effect of graft size on graft fracture rate after anterior lumbar spinal fusion in a sheep model.

Kubosch David, Milz Stefan, Sprecher Christoph M, Südkamp Norbert P, Müller Christof A, Strohm Peter C

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

Sheep anterior spinal fusion model with autogenous and xenogenous bone grafts.

OBJECTIVE

To investigate whether the relationship between cross-sectional area of the bone graft and area of the adjacent vertebral endplates has an effect on graft fracture rate.

SUMMARY OF BACKGROUND DATA: Anterior spondylodesis with autogenous iliac crest transplant is a frequently performed operation to stabilize spinal motion segments but to date no precise recommendations with respect to minimum graft size are available in the literature.

METHODS

Anterior spondylodesis using autogenous and xenogenous grafts of constant size in combination with an angular stable plate (Macs TL). Autogenous iliac crest graft was inserted in eight sheep and xenogenic, commercially available bovine graft (Tutobone) in the additional eight animals. The surface areas of the endplates of the fused intervertebral space were calculated using CT scans and contact radiographs of the specimens obtained after 24 weeks. The graft itself was evaluated for fractures and osteolysis.

RESULTS

A fracture occurred in tricortical, autogenous grafts if the graft cross-sectional area was less than 21% of the area of the adjacent endplates. All xenogenic grafts fractured and therefore a comparable value could not be determined.

CONCLUSION

The results clearly indicate that the relation between graft cross-sectional area and endplate area defines the survival or fracture of the graft in anterior spinal fusion. Although it is difficult to directly apply the results to the clinical situation it is suggested to choose a sufficiently large graft, in order to reduce the risk of autogenous graft fracture in anterior spondylodesis.

Для цитирования: Kubosch David, Milz Stefan, Sprecher Christoph M, Südkamp Norbert P, Müller Christof A, Strohm Peter C. Effect of graft size on graft fracture rate after anterior lumbar spinal fusion in a sheep model.. Injury. 2009. DOI: 10.1016/j.injury.2009.08.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.