MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/euy304

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2019

Omnipolarity applied to equi-spaced electrode array for ventricular tachycardia substrate mapping.

Porta-Sánchez Andreu, Magtibay Karl, Nayyar Sachin, Bhaskaran Abhishek, Lai Patrick F H, Massé Stéphane, Labos Christopher, Qiang Beiping, Romagnuolo Rocco, Masoudpour Hassan, Biswas Labonny, Ghugre Nilesh, Laflamme Michael, Deno Don Curtis, Nanthakumar Kumaraswamy

СтатьяПеревод в очереди

Аннотация

AIMS: Bipolar electrogram (BiEGM)-based substrate maps are heavily influenced by direction of a wavefront to the mapping bipole. In this study, we evaluate high-resolution, orientation-independent peak-to-peak voltage (Vpp) maps obtained with an equi-spaced electrode array and omnipolar EGMs (OTEGMs), measure its beat-to-beat consistency, and assess its ability to delineate diseased areas within the myocardium compared against traditional BiEGMs on two orientations: along (AL) and across (AC) array splines.

METHODS AND

RESULTS

The endocardium of the left ventricle of 10 pigs (three healthy and seven infarcted) were each mapped using an Advisor™ HD grid with a research EnSite Precision™ system. Cardiac magnetic resonance images with late gadolinium enhancement were registered with electroanatomical maps and were used for gross scar delineation. Over healthy areas, OTEGM Vpp values are larger than AL bipoles by 27% and AC bipoles by 26%, and over infarcted areas OTEGM Vpp values are 23% larger than AL bipoles and 27% larger than AC bipoles (P < 0.05). Omnipolar EGM voltage maps were 37% denser than BiEGM maps. In addition, OTEGM Vpp values are more consistent than bipolar Vpps showing less beat-by-beat variation than BiEGM by 39% and 47% over both infarcted and healthy areas, respectively (P < 0.01). Omnipolar EGM better delineate infarcted areas than traditional BiEGMs from both orientations.

CONCLUSION

An equi-spaced electrode grid when combined with omnipolar methodology yielded the largest detectable bipolar-like voltage and is void of directional influences, providing reliable voltage assessment within infarcted and non-infarcted regions of the heart.

Для цитирования: Porta-Sánchez Andreu, Magtibay Karl, Nayyar Sachin, Bhaskaran Abhishek, Lai Patrick F H, Massé Stéphane, Labos Christopher, Qiang Beiping, Romagnuolo Rocco, Masoudpour Hassan, Biswas Labonny, Ghugre Nilesh, Laflamme Michael, Deno Don Curtis, Nanthakumar Kumaraswamy. Omnipolarity applied to equi-spaced electrode array for ventricular tachycardia substrate mapping.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2019. DOI: 10.1093/europace/euy304.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.