DOI 10.1016/j.knee.2016.10.011
The Genovese grading scale is not reliable for MR assessment of collagen meniscus implants.
Аннотация
BACKGROUND
The purpose of the study was to evaluate the intra- and inter-observer reliabilities of the Genovese grading on MRI in patients after collagen meniscus substitution.
METHODS
84 MRI images of 74 consecutive patients who underwent partial meniscus substitution using collagen meniscus implant (CMI) were assessed. MRIs were evaluated using the Genovese grading system. Furthermore, meniscal extrusion was assessed. Two observers performed the grading twice, blinded to each other and to the previous results, with a six weeks interval. The inter- and intra-observer reliabilities were assessed using kappa and weighted kappa values.
RESULTS
The criterion "morphology/size" showed a weighted kappa value inter-observer reliability of 0.069 (rater 1)/0.352 (rater 2) and intra-observer reliability of 0.170 (1st rating)/0.582 (2nd rating). The criterion "signal intensity" showed a weighted kappa value inter-observer reliability of 0.175/0.284 and intra-observer reliability of 0.294/0.458. The criterion "cartilage lesions" showed a kappa value inter-observer reliability of 0.091/0.525 and intra-observer reliability of 0.409/0.413. The criterion "bone marrow edema" showed a kappa value inter-observer reliability of 0.667/0.808 and intra-observer reliability of 0.702/0.715. The criterion "cartilage lesions" showed a kappa value inter-observer reliability of 0.091/0.525 and intra-observer reliability of 0.409/0.413. Regarding meniscal extrusion kappa values for the inter-observer reliability were 0.625/0.940 and 0.625/0.889 for intra-observer reliability.
CONCLUSIONS
Three of the four Genovese grading items showed only slight to moderate inter- and intra-observer reliabilities in evaluating CMI on MRI. Hence, such grading results need to be considered with all due care. Only the criteria "bone marrow edema" and "meniscal extrusion" showed a good agreement for both inter- and intra-observer reliabilities.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.