Te Moller N C R, Pitkänen M, Sarin J K, Väänänen S, Liukkonen J, Afara I O, Puhakka P H, Brommer H, Niemelä T, Tulamo R-M, Argüelles Capilla D, Töyräs J
Аннотация
BACKGROUND: Arthroscopic optical coherence tomography (OCT) is a promising tool for the detailed evaluation of articular cartilage injuries. However, OCT-based articular cartilage scoring still relies on the operator's visual estimation.
OBJECTIVES: To test the hypothesis that semi-automated International Cartilage Repair Society (ICRS) scoring of chondral lesions seen in OCT images could enhance intra- and interobserver agreement of scoring and its accuracy.
STUDY DESIGN: Validation study using equine cadaver tissue.
METHODS: Osteochondral samples (n = 99) were prepared from 18 equine metacarpophalangeal joints and imaged using OCT. Custom-made software was developed for semi-automated ICRS scoring of cartilage lesions on OCT images. Scoring was performed visually and semi-automatically by five observers, and levels of inter- and intraobserver agreement were calculated. Subsequently, OCT-based scores were compared with ICRS scores based on light microscopy images of the histological sections of matching locations (n = 82).
RESULTS: When semi-automated scoring of the OCT images was performed by multiple observers, mean levels of intraobserver and interobserver agreement were higher than those achieved with visual OCT scoring (83% vs. 77% and 74% vs. 33%, respectively). Histology-based scores from matching regions of interest agreed better with visual OCT-based scoring than with semi-automated OCT scoring; however, the accuracy of the software was improved by optimising the threshold combinations used to determine the ICRS score.
MAIN LIMITATIONS: Images were obtained from cadavers.
CONCLUSIONS: Semi-automated scoring software improved the reproducibility of ICRS scoring of chondral lesions in OCT images and made scoring less observer-dependent. The image analysis and segmentation techniques adopted in this study warrant further optimisation to achieve better accuracy with semi-automated ICRS scoring. In addition, studies on in vivo applications are required.
OBJECTIVES: To test the hypothesis that semi-automated International Cartilage Repair Society (ICRS) scoring of chondral lesions seen in OCT images could enhance intra- and interobserver agreement of scoring and its accuracy.
STUDY DESIGN: Validation study using equine cadaver tissue.
METHODS: Osteochondral samples (n = 99) were prepared from 18 equine metacarpophalangeal joints and imaged using OCT. Custom-made software was developed for semi-automated ICRS scoring of cartilage lesions on OCT images. Scoring was performed visually and semi-automatically by five observers, and levels of inter- and intraobserver agreement were calculated. Subsequently, OCT-based scores were compared with ICRS scores based on light microscopy images of the histological sections of matching locations (n = 82).
RESULTS: When semi-automated scoring of the OCT images was performed by multiple observers, mean levels of intraobserver and interobserver agreement were higher than those achieved with visual OCT scoring (83% vs. 77% and 74% vs. 33%, respectively). Histology-based scores from matching regions of interest agreed better with visual OCT-based scoring than with semi-automated OCT scoring; however, the accuracy of the software was improved by optimising the threshold combinations used to determine the ICRS score.
MAIN LIMITATIONS: Images were obtained from cadavers.
CONCLUSIONS: Semi-automated scoring software improved the reproducibility of ICRS scoring of chondral lesions in OCT images and made scoring less observer-dependent. The image analysis and segmentation techniques adopted in this study warrant further optimisation to achieve better accuracy with semi-automated ICRS scoring. In addition, studies on in vivo applications are required.