Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study.

Petre Benjamin M, Smith Sean D, Jansson Kyle S, de Meijer Peter-Paul, Hackett Thomas R, LaPrade Robert F, Wijdicks Coen A

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Optimization of anterior cruciate ligament (ACL) fixation is desired to improve graft healing. New soft tissue cortical suspension devices for femoral tunnel fixation should be biomechanically evaluated.

HYPOTHESIS: All femoral fixation devices would prevent a clinically significant amount of displacement and support loads significantly larger than in situ forces experienced by the ACL during early rehabilitation.

STUDY

DESIGN

Controlled laboratory study.

METHODS

Four cortical soft tissue ACL graft suspension devices were tested under cyclic and pull-to-failure loading conditions in both an isolated device-only setup and as a complete bone-device-tendon construct in porcine femurs using a tensile testing machine.

RESULTS

There were significant differences in the ultimate failure loads among the devices. The highest ultimate failure loads when tested as a construct were observed for the XO Button (1748 N), followed by the Endobutton CL (1456 N), ToggleLoc with ZipLoop (1334 N), and TightRope RT (859 N). Cyclic displacement after 1000 cycles during isolated device testing was less than 1 mm for all devices. Cyclic displacements after 1000 cycles in the porcine construct were 1.88 mm, 2.74 mm, 3.34 mm, and 1.82 mm for the Endobutton, TightRope, ToggleLoc, and XO Button, respectively; all were significantly different from each other except when the Endobutton was compared with the XO Button. The ToggleLoc exceeded the 3.0-mm displacement threshold defined as a clinical failure. The most displacement occurred during the first cycle, especially for the adjustable-length loop devices. Stiffness reapproximated the native ACL stiffness for all constructs.

CONCLUSION

The Endobutton, TightRope, and XO Button have the necessary biomechanical properties with regard to ultimate failure strength, displacement, and stiffness for initial fixation of soft tissue grafts in the femoral tunnel for ACL reconstruction. The ToggleLoc had sufficient ultimate failure strength but crossed our 3.0-mm clinical failure threshold for cyclic displacement. Although this study was not designed to compare fixed and adjustable-length loop devices, it was noted that both fixed-loop devices allowed less cyclic displacement and initial displacement.

CLINICAL RELEVANCE: Adjustable-length loop devices may need to be retensioned after cycling the knee and fixing the tibial side to account for the increased initial displacement seen with these devices.

Для цитирования: Petre Benjamin M, Smith Sean D, Jansson Kyle S, de Meijer Peter-Paul, Hackett Thomas R, LaPrade Robert F, Wijdicks Coen A. Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study.. The American journal of sports medicine. 2012. DOI: 10.1177/0363546512469875.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.