DOI 10.1016/j.jhsa.2021.03.014
A Prospective Randomized Comparison of Variable-Angle and Fixed-Angle Volar Locking Plating for Intra-Articular Distal Radius Fractures.
Аннотация
PURPOSE: To compare clinical and radiographic outcomes of using a variable-angle volar locking plate (VAVLP) with those of using a fixed-angle volar locking plate (FAVLP) for treating unstable intra-articular fractures of the distal radius.
METHODS
One hundred twenty patients with unstable intra-articular fractures of the distal radius were randomized to open reduction and internal fixation with a VAVLP (n = 60) or an FAVLP (n = 60). Supplementary methods (eg., Kirschner wire fixation) were required in 4 patients with a VAVLP and 9 with an FAVLP. Clinical outcomes were evaluated at 6 weeks, 3 months, 6 months, and 1 year after surgery. Posteroanterior and lateral radiographs were used to measure standard radiographic parameters before surgery, in the immediate postoperative period, and at 1 year. Plate prominence and articular congruity were quantified using computed tomography at 6 months.
RESULTS
There were no significant differences in any clinical outcome between the groups at any follow-up time. Volar tilt was significantly greater in patients treated with a FAVLP in the immediate postoperative period (8° vs 6°) and at 1 year (8° vs 5°). Although significant differences were not found in articular gap or stepoff between the 2 plates, the distal and volar prominence of the VAVLP was significantly greater than that of the FAVLP at 6 months. Significantly more patients treated with a VAVLP had a complication (38% vs 19%). However, most secondary surgeries were performed for hardware removal, and no patients from either group had complex regional pain syndrome or tendon rupture.
CONCLUSIONS
Patients with intra-articular distal radius fractures can expect good functional and radiographic outcomes with VAVLP or FAVLP fixation. The VAVLP may be more prone to technical errors, leading to complications, whereas the FAVLP is more likely to require supplementary fixation.
TYPE OF STUDY/LEVEL OF EVIDENCE: Therapeutic I.