Does Provisional Minifragment Fixation Prevent Compression With Dynamic Compression Plating? A Biomechanical Analysis.

Wheatley Benjamin M, Schimoler Patrick J, Hempen Eric C, Miller Mark Carl, Westrick Edward, Altman Gregory T

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To compare the compressive force generated by a 3.5-mm compression plate with and without provisional fixation using a 2.0-mm minifragment plate.

METHODS

Fourth generation composite large humeral sawbones underwent transection and were divided into 2 groups. The first group underwent fixation with a 3.5-mm compression plate; the second group underwent provisional fixation with a 2.0-mm plate followed by definitive fixation using a 3.5-mm plate. Using a load cell, the compressive force generated was measured after insertion of each of 2 eccentrical placed screws and the total compression recorded.

RESULTS

There was no difference in the force generated after each successive compression screw (P = 0.59 and 0.58, respectively). Likewise, there was no significant difference in the total compression generated when the preload was accounted for (P = 0.93).

CONCLUSION

Provisional minifragment fixation does not have any adverse effect on the forces generated during compression plating. These findings suggest that provisional minifragment plates do not need to be removed before definitive fixation.

Для цитирования: Wheatley Benjamin M, Schimoler Patrick J, Hempen Eric C, Miller Mark Carl, Westrick Edward, Altman Gregory T. Does Provisional Minifragment Fixation Prevent Compression With Dynamic Compression Plating? A Biomechanical Analysis.. Journal of orthopaedic trauma. 2021. DOI: 10.1097/bot.0000000000002059.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.