MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/jor.20701

Медицинская библиотека · Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2008

Serum ion levels after ceramic-on-ceramic and metal-on-metal total hip arthroplasty: 8-year minimum follow-up.

Savarino Lucia, Padovani Giovanni, Ferretti Massimo, Greco Michelina, Cenni Elisabetta, Perrone Giorgio, Greco Francesco, Baldini Nicola, Giunti Armando

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Alternative bearing surfaces for total hip arthroplasty, such as metal-on-metal and ceramic-on-ceramic, offer the potential to reduce mechanical wear and osteolysis. In the short and medium term, the second generation of metal-on-metal bearings demonstrated high systemic metal ion levels, whereas ceramic-on-ceramic bearings showed the lowest ones. We aimed to verify whether the long-term ion release in metal-on-metal subjects was still relevant at a median 10-year follow-up, and whether a fretting process at the modular junctions occurred in ceramic-on-ceramic patients and induced an ion dissemination. Serum levels were measured in 32 patients with alumina-on-alumina implants (group A), in 16 subjects with metal-on-metal implants (group B), and in 47 healthy subjects (group C). Group B results were compared with medium-term findings. Cobalt and chromium levels were significantly higher in metal-on-metal implants than in ceramic-on-ceramic ones and controls. Nevertheless, ion levels showed a tendency to decrease in comparison with medium-term content. In ceramic-on-ceramic implants, ion values were not significantly different from controls. Both in groups A and B, aluminum and titanium release were not significantly different from controls. In conclusion, negligible serum metal ion content was revealed in ceramic-on-ceramic patients. On the contrary, due to the higher ion release, metal-on-metal coupling must be prudently considered, especially in young patients, in order to obtain definitive conclusions.

Для цитирования: Savarino Lucia, Padovani Giovanni, Ferretti Massimo, Greco Michelina, Cenni Elisabetta, Perrone Giorgio, Greco Francesco, Baldini Nicola, Giunti Armando. Serum ion levels after ceramic-on-ceramic and metal-on-metal total hip arthroplasty: 8-year minimum follow-up.. Journal of orthopaedic research : official publication of the Orthopaedic Research Society. 2008. DOI: 10.1002/jor.20701.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.