Acetabular component orientation in intra- and postoperative positions in total hip arthroplasty.
Аннотация
INTRODUCTION
In total hip arthroplasty (THA), acetabular component orientation has critically important effects on dislocation, range of motion, polyethylene wear, pelvic osteolysis, and component migration. The differences in the pelvic orientation in the intraoperative lateral position for insertion of acetabular component during operation and that in the postoperative supine position for evaluation of acetabular component orientation will be one of the factors, which make outliers in acetabular component orientation. We compared acetabular component orientation between intraoperative lateral position and postoperative supine position in 100 consecutive primary THAs.
MATERIALS AND
METHODS
A total of 100 consecutive primary THAs (between October 2004 and December 2005) in 100 patients performed by a single surgical team were investigated. Intraoperative anteroposterior radiographs of pelvis in the lateral position and postoperative anteroposterior radiographs of pelvis in the supine position were taken. Acetabular component orientation (vertical tilt and anteversion) were measured using computer software.
RESULTS
The absolute values of difference between measurements in the two positions were 5.3 degrees +/- 4.5 degrees (mean +/- SD) for vertical tilt and 5.1 degrees +/- 3.7 degrees for anteversion. The difference in the vertical tilt between the two positions was significant (P < 0.0001).
CONCLUSION
The difference in the acetabular component orientation between the two positions, which might be caused by the difference between intra- and postoperative pelvic orientation, should be considered during THA.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.