DOI 10.1002/jbm.b.31048

Медицинская библиотека · Journal of biomedical materials research. Part B, Applied biomaterials · 2008

Wear particle analysis of highly crosslinked polyethylene isolated from a failed total hip arthroplasty.

Minoda Yukihide, Kobayashi Akio, Sakawa Akira, Aihara Masaharu, Tada Koichi, Sugama Ryo, Iwakiri Kentaro, Ohashi Hirotsugu, Takaoka Kunio

СтатьяПеревод в очереди

Аннотация

Polyethylene wear particles are one of the most important factors affecting the results of total hip arthroplasty (THA). To reduce wear generation and to achieve better long-term results of THA, highly crosslinked polyethylene (HXPE) has recently been introduced and come into wide use. Thus far, however, there have been no reports on in vivo analysis of HXPE wear particles. We isolated HXPE wear particles from periprosthetic tissue of a failed THA and analyzed using scanning electron microscope. The number of particles was 5.33 x 10(7) g(-1). Particle size (equivalent circle diameter) was 0.66 +/- 0.40 microm (mean +/- standard error). Aspect ratio and roundness were 1.37 +/- 0.26 and 1.44 +/- 0.67, respectively. All the particles were round shaped, and "fibrils" or "shreds" were not detected. Thus far, this was the first report on in vivo wear particle analysis of HXPE. HXPE generated less, smaller, and rounder particles, compared with the corresponding reported values for particles generated from conventional polyethylene. These characteristics might affect macrophage response, osteolysis, and long-term results of THA with HXPE.

Для цитирования: Minoda Yukihide, Kobayashi Akio, Sakawa Akira, Aihara Masaharu, Tada Koichi, Sugama Ryo, Iwakiri Kentaro, Ohashi Hirotsugu, Takaoka Kunio. Wear particle analysis of highly crosslinked polyethylene isolated from a failed total hip arthroplasty.. Journal of biomedical materials research. Part B, Applied biomaterials. 2008. DOI: 10.1002/jbm.b.31048.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.