Highly cross-linked polyethylene bearing surfaces in total hip arthroplasty.
Аннотация
Polyethylene wear-induced osteolysis is the most significant primary factor limiting the life span of total joint arthroplasty. To reduce ultra-high-molecular-weight polyethylene (UHMWPE) particulate wear debris, highly cross-linked polyethylene (HXPE) bearings have been introduced in total hip arthroplasty (THA). In vitro hip simulator wear studies with HXPE have demonstrated a decrease in volumetric wear at the hip by 42% to 100% when compared with conventional metal-on-polyethylene bearings. Early to intermediate clinical results suggest that the in vivo wear properties of HXPE products are superior to those of conventional UHMWPE. Second-generation HXPE materials that utilize alternate cross-linking and free radical quenching techniques have been developed and propose to further minimize wear and oxidation.