[Relationship between Radiation Quality and Image Quality in Digital Chest Radiography: Validation Study Using Human Soft Tissue-equivalent Phantom].

Kawashima Hiroki, Ichikawa Katsuhiro, Kunitomo Hiroshi

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate image quality for chest radiography at different radiation qualities, using phantoms with scatter fractions similar to those of lungs.

METHODS

Two base phantoms with 10 and 4 cm thicknesses, respectively, made of a soft tissue-equivalent material, were used to mimic the X-ray attenuation of the human lung. Two plates with soft tissue- and bone-equivalent materials, respectively, were placed on the base phantom as contrast objects. The image data were obtained with the same entrance surface dose in each radiation quality. Six radiation qualities generated using 120 and 90 kV, and additional copper filters with thicknesses 0, 0.1, and 0.2 mm were selected. The signal-difference-to-noise ratio (SdNR) and a contrast ratio of the soft tissue to the bone were measured for the six radiation qualities.

RESULTS

The thicker the additional filter, the better the SdNR at both tube voltages. The SdNR values were not significantly different between 120 and 90 kV for the same filter thickness. The contrast ratio was higher at 120 than at 90 kV by approximately 8%.

CONCLUSIONS

Because of the advantage of the contrast ratio and the highest SdNR, the radiation quality with 120 kV and 0.2-mm copper filtration was the best. It was indicated that the conventional tube voltage of 120 kV remains to be better than the lower tube voltage of 90 kV.

Для цитирования: Kawashima Hiroki, Ichikawa Katsuhiro, Kunitomo Hiroshi. [Relationship between Radiation Quality and Image Quality in Digital Chest Radiography: Validation Study Using Human Soft Tissue-equivalent Phantom].. Nihon Hoshasen Gijutsu Gakkai zasshi. 2021. DOI: 10.6009/jjrt.2021_jsrt_77.3.255.