Journal of prosthodontics : official journal of the American College of Prosthodontists · 2021

Effect of Crown Height on the Screw Stability of Titanium Screw-Retained Crowns.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Yilmaz Burak, Batak Burcu, Seghi Robert, Johnston William M, Lang Lisa A

Аннотация

PURPOSE: The aim of this in vitro study was to evaluate the effect of crown height on the screw stability of screw-retained titanium implant crowns subjected to cyclic loading conditions.

MATERIALS AND METHODS: Twenty-one implants with internal hex connections were placed in epoxy resin holders. Mandibular first molar screw-retained titanium implant crowns with UCLA type, crown-abutment connections were CAD/CAM fabricated. Seven crowns of 3 different heights (6 mm, 10 mm, and 14 mm) were made. The crowns were seated onto the implants and screws were tightened to 30 Ncm. The implants were clamped into holders and stepwise cyclic loads were applied to the occlusal surface at 30-degree angles to the long axes of the crowns. The detorque values were measured after each 5 million cycles. Before increasing the applied load, the crowns were secured with new screws and tightened to 30 Ncm. Failure times, survival estimates and detorque values were then analyzed. (alpha = 0.05).

RESULTS: Crown height did not significantly affect detorque values. However, five 14-mm crowns failed with varying fractures during the 475 N loading condition. Overall, a significantly lower survival for 14 mm crowns was found compared to 6 mm and 10 mm crowns (p = 0.004).

CONCLUSIONS: Crown heights of one-piece screw-retained titanium implant crowns did not significantly affect detorque values. Screw fracture, however, was greater for crown height of 14 mm than those of 6 mm and 10 mm.