DOI 10.1016/j.medengphy.2021.02.003
Biomechanical optimization of the far cortical locking technique for early healing of distal femur fractures.
Аннотация
This finite element study optimized far cortical locking (FCL) technology for early callus formation in distal femur fracture fixation with a 9-hole plate using FCL screws proximal to, and standard locking screws distal to, the fracture. Analyses were done for 120 possible FCL screw configurations by varying FCL screw distribution and number. A hip joint force of 700 N (i.e. 100% x body weight) was used, which corresponds to a typical 140 N "toe-touch" foot-to-ground force (i.e. 20% x body weight) suggested to patients immediately after surgery. Increased FCL screw distribution (i.e. shorter plate working length) caused a decrease at the medial side and an increase at the lateral side of the axial interfragmentary motion (AIM), mildly affected shaft and condylar cortex Von Mises max stress (σMAX), increased plate σMAX, and decreased shaft FCL screw and condylar locking screw σMAX. Increased FCL screw number decreased AIM and σMAX on the shaft cortex, condylar cortex, plate, and FCL screws, but not condylar screws. The optimal FCL screw configuration had 3 FCL screws in plate holes #1, 5, and 6 (proximal to distal) for optimal AIM of 0.2 - 1 mm and reduce shear fracture motion, thereby encouraging early callus formation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.