DOI 10.1016/j.jhsa.2006.09.009

Медицинская библиотека · The Journal of hand surgery · 2006

Wear-induced osteolysis and synovial swelling in a patient with a metal-polyethylene wrist prosthesis.

Groot Diederik, Gosens Taco, Leeuwen Niels C M V, Rhee Marina V, Teepen Hans J L J M

Клинический случайСтатьяПеревод в очереди

Аннотация

We present a patient with a metal and polyethylene wrist prosthesis who had recurrent synovial swelling on the volar side of the wrist. Surgical exploration showed synovitis, a severely worn and broken polyethylene insert, polyethylene wear particles, and titanium debris from fretting corrosion. Pathological analysis showed macrophage-phagocytosing polyethylene wear particles and titanium oxide debris. This shows that polyethylene and titanium wear in wrist prosthesis might be a bigger problem than originally believed.

Для цитирования: Groot Diederik, Gosens Taco, Leeuwen Niels C M V, Rhee Marina V, Teepen Hans J L J M. Wear-induced osteolysis and synovial swelling in a patient with a metal-polyethylene wrist prosthesis.. The Journal of hand surgery. 2006. DOI: 10.1016/j.jhsa.2006.09.009.