Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments.

Oh Soram, Kum Kee-Yeon, Kim Hyun-Jung, Moon Su-Young, Kim Hyeon-Cheol, Chaniotis Antonis, Perinpanayagam Hiran, Pedullá Eugenio, Chang Seok Woo

СтатьяПеревод в очереди

Аннотация

BACKGROUND/PURPOSE: Several single-file systems manufactured using different heat treatment and operated by different kinematics have been released. This study compared the bending resistance and cyclic fatigue resistance of three NiTi files, and examined their phase-transformation behavior.

MATERIALS AND

METHODS

WaveOne Gold Primary (WOG), Reciproc Blue R25 (RPB), and HyFlex EDM OneFile (HDM) were tested (N = 40/instrument). A vertical load was applied to a point 3 mm from the tip, and the stress was measured until a displacement of 3 mm. Tests were conducted at either room temperature (RT: 22 °C) or body temperature (BT: 37 °C) (n = 10). Cyclic fatigue resistance tests were performed in an artificial canal, which had a curvature angle of 40° and a 5-mm radius. Tests were conducted at either RT or BT (n = 10). Instruments were operated according to the manufacturers' instructions. Test results were analyzed using the Kruskal-Wallis and the Mann-Whitney tests. Additional three instruments of each brand were subjected to differential scanning calorimetry (DSC).

RESULTS

At RT the bending resistance of three files were not significantly different. However, at BT the bending resistance of RPB was highest, followed by WOG, and HDM (P < 0.05). At RT, RPB demonstrated the longest fracture time, followed by HDM, and WOG (P < 0.05). At BT, HDM had the longest fracture time, followed by RPB, and WOG (P < 0.05). The WOG, RPB consisted of austenite in a considerable proportion, whereas HDM was mainly martensite state at BT.

CONCLUSION

HDM presented superior flexibility and cyclic fatigue resistance at BT.

Для цитирования: Oh Soram, Kum Kee-Yeon, Kim Hyun-Jung, Moon Su-Young, Kim Hyeon-Cheol, Chaniotis Antonis, Perinpanayagam Hiran, Pedullá Eugenio, Chang Seok Woo. Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments.. Journal of dental sciences. 2019. DOI: 10.1016/j.jds.2019.10.003.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.