In vitro assessment of PEEK and titanium implant abutments: Screw loosening and microleakage evaluations under dynamic mechanical testing.

Ortega-Martínez Jordi, Delgado Luis M, Ortiz-Hernández Mónica, Punset Miquel, Cano-Batalla Jordi, Cayon Miguel Roig, Cabratosa-Termes Josep

СтатьяПеревод в очереди

Аннотация

STATEMENT OF PROBLEM: Polyetheretherketone (PEEK) has been advocated to replace metal components in dentistry. Although PEEK is a high-performance polymer with a white color, adequate biological response, and resistance to fracture, data to support PEEK as an alternative material for implant abutments are lacking.

PURPOSE: The purpose of this in vitro study was to assess the mechanical and functional properties of PEEK implant abutments as a nonmetallic alternative to titanium abutments, which presents esthetic limitations and greater difficulty to customize clinically.

MATERIAL AND

METHODS

Implant abutments manufactured by using PEEK (n=24) or titanium grade 5 (n=24) were attached to MIS Implants type M4 3.75×16 mm with an internal screw tightened to 25 Ncm. Screw loosening and microleakage was assessed by submersion in a 2% methylene blue solution for 48 hours at 37 °C. Both groups were compared with and without applying dynamic loading; fatigue testing was performed following the International Organization for Standardization (ISO) 14801:2016 standard. All specimens were observed under a stereomicroscope at ×8 magnification. Statistically significant differences among the PEEK and titanium implant abutments were tested with 2-factor ANOVA and the chi-square analysis for nonpaired and paired data, respectively (α=.05).

RESULTS

The implant abutments made of titanium were better in all mechanical tests. The torque loss of titanium abutments was approximately 10%, while PEEK showed a significantly higher (P<.05) torque loss of up to 50%. Moreover, 91.6% of the titanium abutments did not present microleakage, while there was no specimen of PEEK abutments without microleakage, once subjected to dynamic loading (P<.05).

CONCLUSIONS

Titanium implant abutments (Ti6Al4V) were better in all tests performed. However, PEEK abutments may be suitable for long-term interim restorations, especially in the anterior area, in patients without parafunction.

Для цитирования: Ortega-Martínez Jordi, Delgado Luis M, Ortiz-Hernández Mónica, Punset Miquel, Cano-Batalla Jordi, Cayon Miguel Roig, Cabratosa-Termes Josep. In vitro assessment of PEEK and titanium implant abutments: Screw loosening and microleakage evaluations under dynamic mechanical testing.. The Journal of prosthetic dentistry. 2020. DOI: 10.1016/j.prosdent.2020.09.033.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.