DOI 10.3390/ma13235318

Медицинская библиотека · Materials (Basel, Switzerland) · 2020

Early Short-Term Postoperative Mechanical Failures of Current Ceramic-on-Ceramic Bearing Total Hip Arthroplasties.

Fernández-Fairén Mariano, Torres-Perez Ana, Perez Roman, Punset Miquel, Molmeneu Meritxell, Ortiz-Hernández Monica, Manero José María, Gil Javier

СтатьяПеревод в очереди

Аннотация

Although ceramic-on-ceramic (CoC) bearings have been shown to produce the smallest amount of wear volume in vitro as well as in vivo studies when used for total hip arthroplasties (THA), concerns about the failure of these bearing surfaces persist due to early failures observed after short postoperative time. In this study, an exhaustive analysis of the early failure occurred on the new generation of ceramic bearings, consisting of a composite alumina matrix-based material reinforced with yttria-stabilized tetragonal zirconia (Y-TZP) particles, chromium dioxide, and strontium crystals, was performed. For this study, 118 CoC bearings from 117 patients were revised. This article describes a group of mechanical failure CoC-bearing BIOLOX THA hip prosthesis patients without trauma history. The retrieved samples were observed under scanning electron microscopy (SEM), composition was analyzed with energy dispersive X-ray spectroscopy (EDX), and damaged surfaces were analyzed by grazing-incidence X-ray diffraction (GI-XRD) and white light interferometry. In the short term, CoC articulations provided similar mechanical behavior and functional outcome to those in XLPE cases. However, 5% more early mechanical failures cases were observed for the ceramic components. Although the fracture rate of third generation CoC couples is low, the present study shows the need to further improve the third generation of CoC-bearing couples for THA. Despite the improved wear compared to other materials, stress concentrators are sources of initial crack propagation, such as those found in the bore-trunnion areas. Moreover, in view of the evidence observed in this study, the chipping observed was due to the presence of monoclinic phase of the Y-TZP instead of tetragonal, which presents better mechanical properties. The results showed that total safety after receiving a THA is still a goal to be pursued.

Для цитирования: Fernández-Fairén Mariano, Torres-Perez Ana, Perez Roman, Punset Miquel, Molmeneu Meritxell, Ortiz-Hernández Monica, Manero José María, Gil Javier. Early Short-Term Postoperative Mechanical Failures of Current Ceramic-on-Ceramic Bearing Total Hip Arthroplasties.. Materials (Basel, Switzerland). 2020. DOI: 10.3390/ma13235318.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.