Contradictory working length effects in locked plating of the distal and middle femoral fractures-a biomechanical study.

Chao Ching-Kong, Chen You-Lin, Wu Jian-Ming, Lin Chen-Huei, Chuang Ting-Yu, Lin Jinn

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Working length have been reported to affect the plate stress and fixation stiffness. However, the results of previous studies have been controversial. The present study was to determine working length effects on different locations of femoral bone gap.

METHODS

Five composite femurs with wide bone gaps at five levels (G1, 2, 3, 5, and 7), were fixed with locking plates. G1-3, G5 and G7 represented gaps at distal femur, distal-middle femur and middle femur respectively. Strain gauges were applied near the screw holes. The plate-bone constructs were loaded through a hemicylinder on the femoral head with total constraints at the distal femur. The micro-strains, axial stiffness and interfragmentary motions were recorded. Then the locking screws were removed one by one and the tests were re-run. The working length effects were compared and correlated.

FINDINGS: In distal femurs (G1-3), long working length was negatively correlated with the highest strains (r = -0.97, -0.95 and - 0.95, p < 0.01) and axial stiffness (r = -1, -0.96 and -0.99, p < 0.01). In distal-middle femurs (G5), as the working length increased, the highest strain decreased initially and then increased (r = 0.81, p = 0.026) and the axial stiffness decreased (r = -0.98, p < 0.01). In middle femurs (G7), the highest strain and gap motions were much higher than that in the other groups and not significantly correlated with the working length change.

INTERPRETATION: Long working length could reduce the highest plate strain in distal femurs, but had no significant effects in middle femurs. The working length effects were markedly affected by the loading and boundary conditions.

Для цитирования: Chao Ching-Kong, Chen You-Lin, Wu Jian-Ming, Lin Chen-Huei, Chuang Ting-Yu, Lin Jinn. Contradictory working length effects in locked plating of the distal and middle femoral fractures-a biomechanical study.. Clinical biomechanics (Bristol, Avon). 2020. DOI: 10.1016/j.clinbiomech.2020.105198.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.