Finite Element Analysis of Insertion Angle of Absorbable Screws for the Fixation of Radial Head Fractures.

Xu Guang-Ming, Liang Zi-Yang, Li Wei, Yang Zheng-Zhong, Chen Zhi-Bin, Zhang Jie

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To investigate the biomechanical effects of different insertion angles of absorbable screws for the fixation of radial head fractures.

METHODS

The finite element models used to simulate the fractures were created based on CT scans. Two absorbable screws were used to fix and maintain the stability of the fracture, and the angles between the screws were set to 0°, 15°, 30°, 45°, 60°, 75°, and 90°. A downward force of 100 N was applied at the stress point, which was coupled with the surface, and the distal radius was limited to six degrees of freedom. The direction and location of the applied force were the same in each model. The values of the von Mises stress and peak displacements were calculated.

RESULTS

Under the applied load and different screw angles, the maximum von Mises stress in the screws was concentrated on the surface contacting the fracture surfaces. The maximum von Mises equivalent stress in the screw decreased when the angle increased from 0° (19.54 MPa) to 45° (13.11 MPa) and increased when the angle further increased to 90° (24.63 MPa). The peak displacement decreased as the angle increased from 0° (0.19 mm) to 45° (0.15 mm) and increased when the angle further increased to 90° (0.25 mm).

CONCLUSION

The computational stress distribution showed that fixation with absorbable screws is safe for patients. Moreover, the minimum von Mises stress and displacements were generated when the angle between the screws was 45°; hence, this setting should be recommended for Mason type II radial fractures.

Для цитирования: Xu Guang-Ming, Liang Zi-Yang, Li Wei, Yang Zheng-Zhong, Chen Zhi-Bin, Zhang Jie. Finite Element Analysis of Insertion Angle of Absorbable Screws for the Fixation of Radial Head Fractures.. Orthopaedic surgery. 2020. DOI: 10.1111/os.12797.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.