DOI 10.2106/jbjs.c.01308

Медицинская библиотека · The Journal of bone and joint surgery. American volume · 2005

Backside wear of polyethylene tibial inserts: mechanism and magnitude of material loss.

Conditt Michael A, Thompson Matthew T, Usrey Molly M, Ismaily Sabir K, Noble Philip C

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Wear of the underside of modular tibial inserts (backside wear) in total knee replacements has been reported by several authors. Although, for some implant designs, this phenomenon seems to contribute to osteolysis, the actual volume of material lost through wear of the backside surface has not been quantified. This study describes the results of computerized measurements of tibial inserts of one design known to be associated with a high prevalence of backside wear in situ.

METHODS

A series of retrieved total knee components of one design were examined. The duration of implantation of the retrieved components ranged from thirty-six to 146 months. Laser surface profilometry and computer-aided design software were used to develop individual three-dimensional models of each worn, retrieved tibial insert to compare with scanned unused inserts. Volumetric subtraction of both models revealed the material lost because of backside wear.

RESULTS

Worn and unworn areas on the backside surface were easily identified by stereomicroscopy and laser profilometry. The computer reconstructions showed that, in all retrievals, all unworn surfaces on the nonarticulating surface lay in one plane. The average volume (and standard deviation) of the material lost because of backside wear was 925 +/- 637 mm(3) (range, 197 to 2720 mm(3)). On the basis of the time in situ for each implant, the average volumetric wear rate was 138 +/- 95 mm(3)/yr.

CONCLUSIONS

The predicted volume of material removed because of backside wear is substantial and may be sufficient to induce osteolysis. Our results suggest that peg-like protrusions are not generated by the extrusion of polyethylene into screw-holes within the base-plate but by abrasion of the underside of the bearing insert, leaving the protruding pegs as the only remnants of the original surface.

Для цитирования: Conditt Michael A, Thompson Matthew T, Usrey Molly M, Ismaily Sabir K, Noble Philip C. Backside wear of polyethylene tibial inserts: mechanism and magnitude of material loss.. The Journal of bone and joint surgery. American volume. 2005. DOI: 10.2106/jbjs.c.01308.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.