DOI 10.1016/j.arth.2004.09.025
Cementless porous-coated sockets without holes implanted with pure press-fit technique.
Аннотация
We evaluated the use of a pure press-fit technique in 220 porous-coated acetabular components without screw holes at mean follow-up of 6.1 years (4-7.5 years). The socket had a 14 degrees enhanced gradient at the periphery and was underreamed by 1 to 2 mm depending on bone quality. The Harris Hip average score was 44.3 (26-70) preoperatively and 93.1 (66-100) at final follow-up. There were no postoperative gaps in any zone. Nonprogressive radiolucent lines were detected in 4 hips. There was no movement or migration of the acetabular component on the follow-up x-rays. Pelvic osteolysis was not observed. Whereas 5 cups (2.3%) had revisions, none was revised for aseptic loosening. The press-fit technique provides stability in appropriate cases avoiding the theoretical complications related to screws or screw holes.