MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrthm.2016.07.003

Медицинская библиотека · Heart rhythm · 2016

Scar voltage threshold determination using ex vivo magnetic resonance imaging integration in a porcine infarct model: Influence of interelectrode distances and three-dimensional spatial effects of scar.

Tung Roderick, Kim Steve, Yagishita Daigo, Vaseghi Marmar, Ennis Daniel B, Ouadah Sarah, Ajijola Olujimi A, Bradfield Jason S, Mahapatra Srijoy, Finn Paul, Shivkumar Kalyanam

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Studies analyzing optimal voltage thresholds for scar detection with electroanatomic mapping frequently lack a gold standard for comparison.

OBJECTIVE

The purpose of this study was to use a porcine infarct model with ex vivo magnetic resonance imaging (MRI) integration to characterize the relationship between interelectrode spacing and bipolar voltage thresholds and examine the influence of 3-dimensional scar on unipolar voltages.

METHODS

Thirty-two combined endocardial-epicardial electroanatomic maps were created in 8 postinfarct porcine subjects (bipolar 2-mm, 5-mm, and 8-mm interelectrode spacing and unipolar) for comparison with ex vivo MRI. Two thresholds were compared: (1) 95% normal distribution and (2) best fit to MRI. Direct electrogram analysis was performed in regions across from MRI-defined scar and adjacent to scar border zone.

RESULTS

A linear increase in optimal thresholds was observed with wider bipole spacing. The 95% thresholds for scar were lower than MRI-matched thresholds with moderate sensitivity for nontransmural scar (54% endo, 63% epi). Unipolar endocardial scar area exceeded MRI-defined scar, resulting in mismatched false scar in 5 of 8 (63%). Endocardial and epicardial unipolar voltages were lower than normal in regions adjacent and across from scar.

CONCLUSION

Variations in interelectrode spacing necessitate tailored bipolar voltage thresholds to optimize scar detection. Statistical 95% thresholds appear to be conservative and not fully sensitive for the detection of scar defined by high-resolution ex vivo MRI. In the presence of endocardial scar, unipolar mapping to quantitatively characterize epicardial scar may be overly sensitive due to 3-dimensional spatial averaging.

Для цитирования: Tung Roderick, Kim Steve, Yagishita Daigo, Vaseghi Marmar, Ennis Daniel B, Ouadah Sarah, Ajijola Olujimi A, Bradfield Jason S, Mahapatra Srijoy, Finn Paul, Shivkumar Kalyanam. Scar voltage threshold determination using ex vivo magnetic resonance imaging integration in a porcine infarct model: Influence of interelectrode distances and three-dimensional spatial effects of scar.. Heart rhythm. 2016. DOI: 10.1016/j.hrthm.2016.07.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.