Mechanical Evaluation of Titanium Plates for Osteoesynthesis High Neck Condylar Fracture of Mandible.

Zieliński Rafał, Kozakiewicz Marcin, Konieczny Bartłomiej, Krasowski Michał, Okulski Jakub

СтатьяПеревод в очереди

Аннотация

BACKGROUND

In the literature no information about plates for the high-neck mandibular condylar osteosynthesis could be found despite that 30 plate designs have been published. The main course consider the basal condylar or diacapitular fractures. The aim of the study was to test mechanically all available designs (only 4 of 30 was proper) on polyurethane mandibles using an individually designed clamping system.

METHODS

Forces required for a 1 mm displacement of fixed fracture fragments and incidents of screw loosening were recorded.

RESULTS

It has occured that dedicated plates for fixation are much weaker than set of two straight plates (p<0.0001). General observation is the bigger plate and more screws, the better rigid stable osteosynthesis of mandibular condyle, however, there are limitations in plates design for high-neck fractures resulted in restricted operation field.

CONCLUSION

Double straight plates occured to be the best mechanical fixation for high-neck fractures of the mandibular condyle. Maybe other existing plates could be used but only after prebending or that fracture required novel dedicated plates design.

Для цитирования: Zieliński Rafał, Kozakiewicz Marcin, Konieczny Bartłomiej, Krasowski Michał, Okulski Jakub. Mechanical Evaluation of Titanium Plates for Osteoesynthesis High Neck Condylar Fracture of Mandible.. Materials (Basel, Switzerland). 2020. DOI: 10.3390/ma13030592.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.