Drainage of infected kidneys with ureteral stents: does size matter?

Modai Jonathan, Shilo Yaniv, Leibovici Dan, Dror Ishai, Kalchenko Vyacheslav, Berkowitz Brian

СтатьяПеревод в очереди

Аннотация

PURPOSE: The purpose of our study was to evaluate the ability of ureteral stents with different diameters to drain pus that accumulates in an obstructed kidney using an in vitro model.

METHODS

We developed an in vitro model of an obstructed kidney filled with pus. The model included a silicon kidney unit based on computed tomography (CT) data, a 3D printed ureteral stone based on a real extracted ureteral stone, a latex ureter model, a bladder vessel, and a fluid with qualities resembling pus. Identical printed stones were inserted into four ureter models containing stents with varying diameters (4.8F, 6F, 7F, 8F), each of which was connected to the kidney unit and the bladder vessel. The kidney unit was filled with artificial pus to pressures of 30 cmH2O to simulate an infected and obstructed kidney. The obstruction was relieved with stents in place, while artificial urine was pumped into the kidney; pressure in the kidney and remaining pus were measured continuously.

RESULTS

The rate of pressure drop and the final pressure measured in the kidney were unaffected by the diameter of the stent. For all stent diameters, the pressure reached non-obstructed levels within 30 s, final pressure was reached within 90-120 s, and minimal amounts of pus remained in the kidney after 120 min.

CONCLUSIONS

In vitro experiments demonstrate that all stent diameters drain pus-filled, obstructed kidneys with the same efficacy. The common perception that larger diameter tubes are more effective under such circumstances should be re-examined.

Для цитирования: Modai Jonathan, Shilo Yaniv, Leibovici Dan, Dror Ishai, Kalchenko Vyacheslav, Berkowitz Brian. Drainage of infected kidneys with ureteral stents: does size matter?. World journal of urology. 2022. DOI: 10.1007/s00345-022-04070-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.