DOI 10.1080/08820538.2024.2443841
Metallic Intraocular Foreign Bodies Following Torsional Phacoemulsification Surgery.
Аннотация
PURPOSE: This study aimed to determine the rate of retained metallic intraocular foreign bodies (IOFBs) following torsional phacoemulsification cataract surgery by an experienced surgeon and elucidate related surgical factors, particle origin theories from existing literature, and postoperative visual impact.
METHODS
An observational study was conducted on 300 consecutive cases of cataract who underwent torsional phacoemulsification by a single surgeon using two-handed stop-and-chop technique. The incidence of retained IOFBs was observed and correlated with the surgical parameters and the density of cataract.
RESULTS
Retained IOFBs were observed in 6 of 300 (2%) cases operated. These were significantly associated with higher cumulative dissipated energy and the density of cataract, but without impacting final vision.
CONCLUSION
Phacoemulsification retains inherent risks of metallic particle liberation from ultrasonically stressed instrumentation interacting within confined spaces. However, small retained IOFBs appear visually inconsequential over limited follow-up. Advances in technique, engineering design, and surgical quality control hold promise for further enhancing safety of the procedure.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.