Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2015
Suitability of the Zenith p-Branch Standard Fenestrated Endovascular Graft for Treatment of Ruptured Abdominal Aortic Aneurysms.
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kristmundsson Thorarinn, Sveinsson Magnus, Björses Katarina, Törnqvist Per, Dias Nuno
Аннотация
PURPOSE: To evaluate the anatomic suitability of the Zenith pivot branch (p-branch) fenestrated device in ruptured abdominal aortic aneurysms (rAAA).
METHODS: Contrast-enhanced computed tomography (CT) images of 206 patients (mean age 75±8 years; 175 men) with rAAA were evaluated in a dedicated 3-dimensional vascular workstation. All aneurysms found unsuitable for standard infrarenal repair were evaluated for Zenith p-branch suitability according to the Investigational Device Exemption protocol for both device configurations (A, pivot fenestrations at the same level; B, right renal fenestration located more cranially).
RESULTS: The suitability of the p-branch (A or B configuration) for short neck aneurysms (<15 mm; n=89) was 49%; of the 26 different combinations of exclusion criteria, a mismatch between a renal artery takeoff and the positioning of the corresponding fenestration was the most common. For juxta- and pararenal aneurysms (neck length <10 mm; n=66), suitability was 48%. Suitability assessed by target vessel positioning only (excluding all other limiting factors) was 58% for short neck aneurysms (n=52) and 55% for juxta- and pararenal aneurysms (n=36).
CONCLUSION: Approximately half of patients with short neck rAAAs would be suitable for the Zenith p-branch fenestrated device according to the instructions for use. In almost 60%, the pivot fenestrations can accommodate the corresponding target vessels. More studies are needed to confirm these findings.
METHODS: Contrast-enhanced computed tomography (CT) images of 206 patients (mean age 75±8 years; 175 men) with rAAA were evaluated in a dedicated 3-dimensional vascular workstation. All aneurysms found unsuitable for standard infrarenal repair were evaluated for Zenith p-branch suitability according to the Investigational Device Exemption protocol for both device configurations (A, pivot fenestrations at the same level; B, right renal fenestration located more cranially).
RESULTS: The suitability of the p-branch (A or B configuration) for short neck aneurysms (<15 mm; n=89) was 49%; of the 26 different combinations of exclusion criteria, a mismatch between a renal artery takeoff and the positioning of the corresponding fenestration was the most common. For juxta- and pararenal aneurysms (neck length <10 mm; n=66), suitability was 48%. Suitability assessed by target vessel positioning only (excluding all other limiting factors) was 58% for short neck aneurysms (n=52) and 55% for juxta- and pararenal aneurysms (n=36).
CONCLUSION: Approximately half of patients with short neck rAAAs would be suitable for the Zenith p-branch fenestrated device according to the instructions for use. In almost 60%, the pivot fenestrations can accommodate the corresponding target vessels. More studies are needed to confirm these findings.
Предметные рубрики MeSH
AgedAged, 80 and overAortic Aneurysm, AbdominalAortic RuptureAortographyBlood Vessel ProsthesisBlood Vessel Prosthesis ImplantationDatabases, FactualEndovascular ProceduresFeasibility StudiesFemaleHumansImaging, Three-DimensionalMalePatient SelectionPredictive Value of TestsProsthesis DesignRadiographic Image Interpretation, Computer-AssistedStentsTomography, X-Ray ComputedTreatment Outcome