Continuous Neural Monitoring in Endoscopic Thyroidectomy: Feasibility Experimental Study for Transcutaneous Vagal Nerve Stimulation.
Аннотация
Background
Adoption of continuous intraoperative neural monitoring (C-IONM) in endoscopic thyroid surgery is limited due to the uneasy application of C-IONM electrode. Feasibility for transcutaneous vagal nerve stimulation was tested. Materials and
Methods
Vagus nerve (VN) and recurrent laryngeal nerve (RLN) were mapped and stimulated in different neck sites by (1) transcutaneous monopolar intermitted stimulation; (2) prototype for continuous transcutaneous stimulating electrodes (TSEs), that is, suction ball C-IONM electrode remodeled; (3) stimulation subsequently to neck CO2 insufflation; (4) direct stimulation of surgical exposed nerves. Electromyographic (EMG) parameters were compared.
Results
Six anterior neck locations evoked EMG signals of VNs and RLNs. Location Nos. 3 and 6 according to our scheme were over the VNs, confirmed by ultrasonography and subsequent dissection. Other locations did not correspond to nerves sites. Transcutaneous thresholds were higher than surgical exposed ones to produce a consistent and satisfactory EMG response. TSE recorded solely in location No. 5 with amplitude values 256-9 μV. It was not possible to stimulate and monitor the RLN and VN after neck CO2 insufflation.
Conclusion
Transcutaneous C-IONM is unfeasible for endoscopic thyroidectomy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.