MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ajem.2023.09.010

Медицинская библиотека · The American journal of emergency medicine · 2023

Evaluation of esophageal injuries after defibrillation with transesophageal ultrasound probe in the mid-esophagus in pigs.

Krulewitz Neil, Levin Nicholas M, Youngquist Scott, Kelly Christopher, Hoareau Guillaume, Johnson Michael Austin, Ockerse Patrick

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Transesophageal echocardiography (TEE) is becoming increasingly utilized by emergency medicine providers during cardiac arrest. Intra-arrest, TEE confers several benefits including shorter pauses in chest compressions and direct visualization of cardiac compressions. Many ultrasound probe manufacturers recommend against performing defibrillation with the TEE probe in the mid-esophagus for fear of causing esophageal injury or damage to the probe, however no literature exists that has investigated this concern. To assess this, we performed cardiopulmonary resuscitation (CPR) and multiple defibrillations in 8 swine with a TEE probe in place.

METHODS

We performed TEE on 8 adult swine during CPR and performed multiple 200 J defibrillations with the TEE probe in the mid-esophagus. Post-mortem, esophagi were dissected and inspected for evidence of injury.

RESULTS

On macroscopic inspection of 8 esophagi, no evidence of hematoma, thermal injury, or perforation was noted.

CONCLUSION

Our study suggests that performing defibrillation during CPR with a TEE probe in place in the mid-esophagus is likely safe and low risk for significant esophageal injury. This further bolsters the use of TEE in CPR and would enable continuous visualization of cardiac activity without the need to remove the TEE probe for defibrillation.

Для цитирования: Krulewitz Neil, Levin Nicholas M, Youngquist Scott, Kelly Christopher, Hoareau Guillaume, Johnson Michael Austin, Ockerse Patrick. Evaluation of esophageal injuries after defibrillation with transesophageal ultrasound probe in the mid-esophagus in pigs.. The American journal of emergency medicine. 2023. DOI: 10.1016/j.ajem.2023.09.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.