Performance of a handheld point of care ultrasonography to assess IUD position compared to conventional transvaginal ultrasonography.

Araujo K G, Yoshida A, Juliato C R T, Sarian L O, Derchain S

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To compare the performance of the abdominal handheld point-of-care ultrasonography (POCUS) Butterfly-iQ to gold standard transvaginal ultrasonography (US) in identifying the position of intrauterine devices (IUDs) in the hands of a medical doctor specialised in ultrasonography.

METHODS

In this diagnostic accuracy study, a single operator conducted abdominal POCUS followed by conventional transvaginal US. Seventy patients utilising copper or hormonal IUDs were assessed between June 2021 and October 2022. IUDs were categorised as entirely within the uterine cavity or malpositioned. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy were calculated for detecting malpositioned IUDs, with conventional US results serving as the reference standard. Concordance rate and Kappa coefficient were computed to assess the agreement between the two ultrasound modalities.

RESULTS

Among the 70 patients, 46 (65.7%) used copper IUDs, and 24 (34.3%) used hormonal IUDs. Conventional transvaginal US showed IUDs entirely within the uterine cavity in 56 (80%) patients and 14 (20%) IUDs were malpositioned. Of the 14 malpositioned IUDs seen by conventional US, POCUS identified 13 demonstrating a sensitivity of 92.9% (66.1-99.8). Of the 56 IUDs entirely within the uterine cavity shown by conventional US, only two cases were considered malpositioned by POCUS demonstrating a specificity of 96.4% (87.7-99.6). The concordance rate was 95.7%, and the Kappa value was 0.87 in differentiating between IUDs entirely within the uterine cavity and those that were malpositioned.

CONCLUSION

Abdominal POCUS using Butterfly-iQ, when administered by an imaging specialist, exhibited excellent performance in confirming IUDs entirely within the uterine cavity.

Для цитирования: Araujo K G, Yoshida A, Juliato C R T, Sarian L O, Derchain S. Performance of a handheld point of care ultrasonography to assess IUD position compared to conventional transvaginal ultrasonography.. The European journal of contraception & reproductive health care : the official journal of the European Society of Contraception. 2024. DOI: 10.1080/13625187.2024.2315231.