Percutaneous retrieval of embolized atrial leadless pacemakers: Early multicenter experience.

Zaidi Syed M, Supple Gregory, Marchlinski Francis E, D'Souza Benjamin, Berkowitz Eric, Rashba Eric, Katz Michael, Feigenblum David, Schaller Robert D

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Leadless dual-chamber pacemakers extend pacing into the atrium but introduce unique challenges related to fixation and stability. Atrial device embolization remains poorly characterized, with limited data beyond isolated case reports.

OBJECTIVES

To describe the clinical features, procedural characteristics, and outcomes of embolized atrial leadless pacemakers, emphasizing warning signs and percutaneous retrieval strategies.

METHODS

We performed a retrospective multicenter case series of AVEIR AR embolization events across 6 United States electrophysiology centers between October 2024 and August 2025. Clinical, procedural, and imaging data were collected, including implant parameters, timing and location of embolization, and retrieval techniques.

RESULTS

Six embolization cases were identified (mean age, 71 ± 10 years, 50% women). All devices were implanted at the base of the right atrial appendage with reported minimal or absent current of injury (COI) at implantation. Embolization occurred intraprocedurally in 1 case and during the early postprocedural period in 5 cases, most commonly to the pulmonary arteries (5 of 6 cases). All devices embolized with the fixation helix oriented proximally. Percutaneous retrieval was successful in all cases using femoral access, deflectable sheaths, and snare-based techniques, without major complications. Reimplantation in 2 of 6 cases yielded a robust COI and favorable chronic electrical parameters.

CONCLUSION

Atrial leadless pacemaker embolization, although uncommon, appears to occur early after implantation and is associated with suboptimal COI at implant. Despite unfavorable device orientation, percutaneous retrieval using standard electrophysiology tools is highly effective. Careful attention to fixation signals and adjunctive imaging may reduce embolization risk.

Для цитирования: Zaidi Syed M, Supple Gregory, Marchlinski Francis E, D'Souza Benjamin, Berkowitz Eric, Rashba Eric, Katz Michael, Feigenblum David, Schaller Robert D. Percutaneous retrieval of embolized atrial leadless pacemakers: Early multicenter experience.. Heart rhythm. 2026. DOI: 10.1016/j.hrthm.2026.07.025.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.