DOI 10.1002/hed.25377

Медицинская библиотека · Head & neck · 2018

Noninvasive, tube-based, continuous vagal nerve monitoring using the laryngeal adductor reflex: Feasibility study of 134 nerves at risk.

Sinclair Catherine F, Téllez Maria J, Ulkatan Sedat

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Continuous vagal intraoperative neuromonitoring (IONM) currently requires placement of a vagal nerve electrode. Herein, we present data from 100 patients (134 nerves-at-risk) monitored continuously during neck endocrine surgeries using a noninvasive, new methodology that solely utilizes endotracheal tube electrodes to simultaneously stimulate laryngeal mucosa and record a laryngeal adductor reflex continuous IONM (LAR-C-IONM) response.

METHODS

The laryngeal adductor reflex (LAR) was elicited by electrical laryngeal mucosal stimulation on the side contralateral to the operative field using endotracheal tube electrodes. All patients completed preoperative and postoperative laryngeal and voice examinations.

RESULTS

One hundred patients (134 nerves-at-risk) were included. Significantly more nerves-at-risk with an LAR opening to closing amplitude decrement >60% or with absolute closing amplitude <100 μV had postoperative vocal fold paralysis (P < .001). The LAR-C-IONM was highly sensitive to recurrent laryngeal nerve (RLN) stretch or compression.

CONCLUSION

The LAR-C-IONM is a promising new way to perform continuous vagal monitoring that requires no equipment other than an electromyography (EMG) endotracheal tube and is undergoing further, large-scale evaluation.

Для цитирования: Sinclair Catherine F, Téllez Maria J, Ulkatan Sedat. Noninvasive, tube-based, continuous vagal nerve monitoring using the laryngeal adductor reflex: Feasibility study of 134 nerves at risk.. Head & neck. 2018. DOI: 10.1002/hed.25377.