The Journal of hand surgery · 2024

A Biomechanical Analysis of Oblique Metacarpal Metadiaphyseal Fracture Fixation in a Cadaver Model.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Garcia-Lopez Edgar, Siu Jeremy Wafong, Kwong Jeffrey W, Kamal Robin N, Schroeder Nicole, Shapiro Lauren M

Аннотация

PURPOSE: Although metacarpal fractures are typically managed nonoperatively, when surgical management is indicated, metacarpal fractures are commonly treated with crossed Kirschner wires (K-wires), which may limit early range of motion. Intramedullary implants are increasing in use with the potential advantage of early range of motion; however, stability in oblique metacarpal neck fractures remains a theoretical concern. The purpose of this study was to determine the biomechanical stability of noncompressive intramedullary fixation for oblique metacarpal neck fractures compared with crossed K-wire fixation.

METHODS: The index, long, and small metacarpals were harvested from three matched pairs of fresh-frozen cadavers. Oblique fractures at the metadiaphyseal region were created in each metacarpal. Each metacarpal was randomized to noncompressive, threaded intramedullary nail fixation or fixation with two crossed K-wires. Specimens were mounted in a Materials Testing System load frame and axially loaded until failure. Load to failure (LTF), stiffness, and load to 2 mm displacement were calculated from load-displacement curves. Differences in peak LTF, stiffness, and load to 2 mm displacement between noncompressive intramedullary fixation and crossed K-wire fixation were evaluated.

RESULTS: The noncompressive intramedullary fixation cohort had a significantly higher LTF (1,190.9 ± 534.7 N vs 297.0 ± 156.0 N) and stiffness (551.3 ± 164.6 N/mm vs 283.0 ± 194.5 N/mm) when compared with the crossed K-wire fixation cohort. Load at 2 mm displacement was greater in the noncompressive intramedullary fixation cohort compared with crossed K-wire fixation (820.5 ± 203.9 vs 514.1 ± 259.6).

CONCLUSIONS: For oblique metadiaphyseal metacarpal fractures, noncompressive intramedullary fixation provides a biomechanically superior construct under axial loading in terms of LTF, stiffness, and load to 2 mm of displacement compared with crossed K-wire fixation.

CLINICAL RELEVANCE: Noncompressive intramedullary nails may be an alternative to K-wire fixation for the treatment of oblique metadiaphyseal metacarpal fractures.