Retrieval of Chronically Implanted Dual-chamber Leadless Pacemakers in an Ovine Model.
Аннотация
BACKGROUND
The clinical utilization of leadless pacemakers (LPs) as an alternative to traditional transvenous pacemakers is likely to increase with the advent of dual-chamber LP systems. Since device retrieval to allow LP upgrade or replacement will become an important capability, the first such dual-chamber, helix-fixation LP system (Aveir DR; Abbott, Abbott Park, IL) was specifically designed to allow catheter-based retrieval. In this study, the preclinical performance and safety of retrieving chronically implanted dual-chamber LPs was evaluated.
METHODS
Atrial and ventricular LPs were implanted in the right atrial appendage and right ventricular apex of 9 healthy ovine subjects. After ≈2 years, the LPs were retrieved using a dedicated transvenous retrieval catheter (Aveir Retrieval Catheter; Abbott) by snaring, docking, and unscrewing from the myocardium. Comprehensive necropsy/histopathology studies were conducted to evaluate device- and procedure-related outcomes.
RESULTS
At a median of 1.9 years postimplant (range, 1.8-2.6), all 18 of 18 (100%) LPs were retrieved from 9 ovine subjects without complications. The median retrieval procedure duration for both LPs, from first-catheter-in to last-catheter-out, was 13.3 minutes (range, 2.5-36.4). Postretrieval, all right atrial, and right ventricular implant sites demonstrated minimal tissue disruption, with intact fibrous tissue limited to the distal device body. No significant device-related trauma, perforation, pericardial effusion, right heart or tricuspid valve injury, or chronic pulmonary thromboembolism were observed at necropsy.
CONCLUSIONS
This preclinical study demonstrated the safe and effective retrieval of chronically implanted, helix-fixation, dual-chamber LP systems, paving the way for clinical studies of LP retrieval.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.