DOI 10.12659/ajcr.940291

Медицинская библиотека · The American journal of case reports · 2023

Delayed Right Ventricular Pacemaker Lead Perforation 9 Years After Implantation.

Hamada Yusuke, Sakaki Masayuki, Watanabe Yoshiki, Hata Satoshi, Kimura Keizo, Sakagoshi Nobuo

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND Cardiac perforation is a rare complication of cardiac implantable electronic devices, with a reported incidence ranging from 0.1% to 5.2%. Delayed perforation, defined as perforation occurring more than 1 month after implantation, is more uncommon. In this report, we present a case of cardiac perforation involving the right ventricle wall that occurred 9 years after pacemaker implantation. CASE REPORT A 79-year-old woman presented with symptoms of dyspnea and was subsequently admitted to a hospital. She had undergone pacemaker implantation for a complete atrioventricular block 9 years prior to the presentation. The patient had right ventricular failure to capture and a resultant complete atrioventricular block. Computed tomography imaging revealed that the right ventricular lead had clearly protruded outside the heart; however, no pericardial effusion was observed. When the patient underwent open surgical repair, the ventricular tined lead was noted to be traversing the right ventricular apex. Device interrogation revealed a sudden increase followed by a gradual decrease in the right ventricular pacing threshold over the course of 2 months, indicating that the lead had slowly traversed the right ventricular muscle before ultimately rupturing through it. CONCLUSIONS This study described the case of a delayed right ventricular pacemaker lead perforation that occurred 9 years after implantation, which was managed via open surgical repair.

Для цитирования: Hamada Yusuke, Sakaki Masayuki, Watanabe Yoshiki, Hata Satoshi, Kimura Keizo, Sakagoshi Nobuo. Delayed Right Ventricular Pacemaker Lead Perforation 9 Years After Implantation.. The American journal of case reports. 2023. DOI: 10.12659/ajcr.940291.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.