Device-related risk of fire in oropharyngeal surgery: a mechanical model.

Roy Soham, Smith Lee P

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

The aim of this study was to evaluate the fire risk for an electrosurgical device (Bovie) as compared with a bipolar radiofrequency ablation wand (Coblator) in a mechanical model of oropharyngeal surgery.

METHODS

An endotracheal tube was inserted into the cranial end of a degutted, whole raw chicken through which 100% oxygen was piped at 10 L/min. An electrosurgical device (15 W, coagulate mode) and bipolar radiofrequency ablation wand (9, 7, and 3 in ablate mode; 5 and 3 in coagulate mode) were activated in the central cavity. All experimental conditions were tested for 4 minutes, or until a positive result was achieved. All trials were repeated to ensure accuracy.

RESULTS

Ignition was obtained with a sustained fire when using the electrosurgical device for between 25 and 80 seconds. Under all 5 experimental conditions, no ignition or sustained fire could be produced using the radiofrequency ablation wand. After 20 sustained minutes of continuous plasma ablation in the chicken cavity without ignition, the electrosurgical device was able to ignite a fire in 25 seconds in the same cavity.

CONCLUSIONS

Although electrosurgical devices present a significant risk of fire during open cavity surgery in oxygen-enriched environments, that risk seems to be eliminated with bipolar radiofrequency plasma ablation.

Для цитирования: Roy Soham, Smith Lee P. Device-related risk of fire in oropharyngeal surgery: a mechanical model.. American journal of otolaryngology. 2009. DOI: 10.1016/j.amjoto.2009.05.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.