Corneal endothelium protection provided by ophthalmic viscosurgical devices during phacoemulsification: experimental study in rabbit eyes.

Park Sally S E, Wilkinson Samuel W, Ungricht Emilie L, Trapnell Michael, Nydegger Jacob, Brintz Ben J, Mamalis Nick, Olson Randall J, Werner Liliana

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

PURPOSE: To compare the protection of the corneal endothelium provided by a cohesive and a dispersive ophthalmic viscosurgical device (OVD) against damage from torsional ultrasound and simulated lenticular debris during phacoemulsification.

SETTING

University setting, Salt Lake City, Utah.

DESIGN

Experimental study.

METHODS

15 New Zealand rabbits were included. After incision, each eye received cohesive or dispersive OVD (ProVisc or Viscoat). 10 1.0 mm beads were injected to simulate lenticular debris. The Intrepid Balanced tip (Centurion Ozil handpiece) was inserted into the anterior chamber, and 60% torsional ultrasound with 50 mL/min flow, 70 mm Hg intraocular pressure, and 600 mm Hg vacuum was applied (20 seconds). Cumulative dissipated energy (CDE) was noted. After euthanasia and enucleation, corneas were removed, stained with trypan blue/alizarin red, and photographed (5 ×400 magnification photos and 1 overview photograph from each cornea). ImageJ was used to evaluate cell damage and loss.

RESULTS

CDE was 4.70 ± 0.26 and 4.64 ± 0.10 in the cohesive and dispersive OVD groups, respectively ( P = .8647). The analysis of the ×400 photographs showed that the percentage of intact cells was statistically higher in the dispersive OVD group ( P = .0002), whereas the percentages of damaged and lost cells were statistically higher in the cohesive OVD group ( P = .0002 and .0059, respectively). Overview photographs revealed the presence of residual OVD on the endothelium, especially in the dispersive group.

CONCLUSIONS

In this experimental study using a rabbit model, the corneal endothelium protection provided by the dispersive OVD was superior to that provided by the cohesive OVD.

Для цитирования: Park Sally S E, Wilkinson Samuel W, Ungricht Emilie L, Trapnell Michael, Nydegger Jacob, Brintz Ben J, Mamalis Nick, Olson Randall J, Werner Liliana. Corneal endothelium protection provided by ophthalmic viscosurgical devices during phacoemulsification: experimental study in rabbit eyes.. Journal of cataract and refractive surgery. 2022. DOI: 10.1097/j.jcrs.0000000000001052.