Pedal Pull-Through Rail and Wingman-Assisted Coaxial Push for Entrapped Device Retrieval in Calcified Tibial Arteries.

Miyazaki Akane, Yoshinaga Masataka, Murata Keisuke, Araki Toru, Matsuo Kouki, Ito Takehiro, Sobue Yoshihiro, Muramatsu Takashi, Fujiwara Wakaya, Watanabe Eiichi

Клинический случайСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To describe a bailout technique for retrieving an entrapped, fractured endovascular device fragment in severely calcified below-the-knee arteries when conventional snare retrieval is not feasible.

KEY STEPS: 1) Immediately stop manipulation and maintain the position of the antegrade guidewire holding the fragment (the "Golden Rule"); 2) obtain retrograde access; 3) advance a snare from the retrograde sheath to capture and externalize the existing antegrade guidewire, creating a pull-through rail without wire exchange; 4) advance a Wingman catheter antegradely over the rail; and 5) apply short, coaxial pushes to dislodge the fragment and retrieve it via the retrograde sheath.

POTENTIAL PITFALLS: Retrograde distal puncture and sheath insertion are difficult without a sufficient vascular diameter in either the dorsal foot artery or posterior tibial artery. Avoid loss of the guidewire during procedure and vessel injury.

TAKE-HOME MESSAGE: Never surrender the workhorse wire holding the fragment; converting it into a pull-through rail enables safe coaxial dislodgement.

Для цитирования: Miyazaki Akane, Yoshinaga Masataka, Murata Keisuke, Araki Toru, Matsuo Kouki, Ito Takehiro, Sobue Yoshihiro, Muramatsu Takashi, Fujiwara Wakaya, Watanabe Eiichi. Pedal Pull-Through Rail and Wingman-Assisted Coaxial Push for Entrapped Device Retrieval in Calcified Tibial Arteries.. JACC. Case reports. 2026. DOI: 10.1016/j.jaccas.2026.107255.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.