An Unexpected Complication after Headless Compression Screw Fixation of an Osteochondral Fracture of Patella.
Аннотация
This study evaluated complications associated with implant depth in headless compression screw treatment of an osteochondral fracture associated with a traumatic patellar dislocation in a 21-year-old woman. Computed tomography and X-rays showed one lateral fracture fragment measuring 25 × 16 mm. Osteosynthesis was performed with two headless compression screws. Five months later, the screws were removed because of patella-femoral implant friction. We recommend that the screw heads be embedded to a depth of at least 3 mm below the cartilage surface. Further clinical studies need to examine the variation in cartilage thickness in the fracture fragment.