DOI 10.1097/brs.0b013e3181e9bf30
Revision strategy for posterior extrusion of the CHARITÉ polyethylene core.
Аннотация
STUDY
DESIGN
This is a case report of a posterior extrusion of the polyethylene core from a CHARITÉ arthroplasty. This is the first reported case of posterior dislocation of the polyethylene and the revision strategies used to correct this problem.
OBJECTIVE
To report a novel failure mechanism and revision strategy for CHARITÉ total disc arthroplasty (TDA).
SUMMARY OF BACKGROUND DATA: Case report at a Level 1 tertiary care referral center in the northeastern United States.
METHODS
This is a case report and review of the literature of a patient who sustained posterior dislocation of the polyethylene core from a CHARITÉ TDA several months after the index procedure.
RESULTS
Core dislocation is a known complication of TDA. However, of the known reported dislocations all have been anterior. This case describes the first known occurrence of posterior core dislocation and the revision strategy for this problem.
CONCLUSION
This case report highlights the first known case of a posterior dislocation of a CHARITÉ core. It is likely that altered biomechanical forces generated over time attributed to device failure. An instrumented posterior fusion with removal of the core is what ultimately led to a stable revision construct.