DOI 10.1016/j.amjoto.2024.104233
Soft tissue injury events associated with neural integrity monitoring endotracheal tubes: A MAUDE database analysis.
Аннотация
OBJECTIVE
Neural integrity monitoring (NIM) endotracheal tubes are widely used to provide intraoperative monitoring of the recurrent laryngeal nerve during certain neck surgeries, especially thyroidectomy, in order to reduce the risk of nerve injury and subsequent vocal fold paralysis. The unique design of NIM tubes and the increased technical skill required for correct placement compared to standard endotracheal tubes may increase the risk of upper aerodigestive tract soft tissue injury. This study aims to describe adverse events related to NIM endotracheal tubes.
STUDY
DESIGN
Retrospective cross-sectional study.
SETTING
The US Food and Drug Administration's MAUDE database (2010-2022); (Manufacturer and User Facility Device Experience).
METHODS
The MAUDE database was queried for reports of adverse events that resulted in patient soft tissue injury involving the use of endotracheal tubes approved by the Food and Drug Administration.
RESULTS
There were 28 reported soft tissue injuries, with all events being related to the NIM EMG family of endotracheal tubes manufactured by Medtronic Xomed, Inc. Overall, 24 were categorized as device-related adverse events, and 4 were unspecified in the event description. The most common soft tissue injuries were edema (n = 7) and perforation (n = 7), each accounting for 25 % of adverse events. The second most common injury type was laceration (n = 4), representing 14 % of all adverse events. Overall, 9 patients (32 %) in our cohort required a surgical intervention to treat their injuries, which consisted of 6 tracheotomies and 3 instances of suture repair.
CONCLUSIONS
The most commonly reported types of soft tissue injury included edema and perforation, followed by laceration. Increased awareness of device-related patient injuries associated with NIM endotracheal tubes can be used to better inform surgeons and anesthesiologists during the process of intubation and surgical decision-making.