Leg length and offset in short-stem total hip arthroplasty: is a single offset-implant sufficient to restore the hip rotation centre within a range of 5 mm?

Olk Felix, Bittersohl Bernd, Babisch Jürgen, Mittelstädt Hagen, Jäger Marcus, Krauspe Rüdiger, Zilkens Christoph

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Short-stem total hip arthroplasty (THA) has gained popularity due to its bone-preserving properties and improved physiological load transmission to the proximal femur. Despite these advantages, the ability of short-stem implants to reliably restore leg length and offset remains debated. This study evaluates whether a single offset implant is sufficient for accurate anatomical reconstruction or if multiple offset options are necessary. MATERIAL AND

METHODS

A total of 148 anteroposterior pelvic radiographs of patients scheduled for short-stem THA were analysed using MediCAD® software. Femoral offset and leg length were measured, and the accuracy of anatomical reconstruction was assessed within a 5-mm target range. Comparisons were made between a single offset (130°) implant and a dual-offset system (130° and 119°) using the McNemar-Bowker test.

RESULTS

With a single 130° CCD (Caput-collum-diaphyseal) offset implant, 55.7% (82/148) of cases achieved satisfactory leg length and offset restoration. The use of a dual-offset system improved accuracy to 79.1% (117/148), demonstrating a statistically significant advantage (p < 0.001). DISCUSSION: The study highlights the need for at least two CCD-angle-offset combinations in short-stem THA to address anatomical variability. A dual-offset system enhances accuracy, reduces biomechanical imbalances, and therefore we expect improvements in clinical outcomes, particularly in teaching hospitals where standardization is essential.

Для цитирования: Olk Felix, Bittersohl Bernd, Babisch Jürgen, Mittelstädt Hagen, Jäger Marcus, Krauspe Rüdiger, Zilkens Christoph. Leg length and offset in short-stem total hip arthroplasty: is a single offset-implant sufficient to restore the hip rotation centre within a range of 5 mm?. Archives of orthopaedic and trauma surgery. 2026. DOI: 10.1007/s00402-026-06194-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.