Fracture Gap Closure and Reduction Are Affected by the Orientation of the Headless Compression Screw.

Mickley John P, Lynch Daniel J, Gordon Adam M, Roebke Austin J, Goyal Kanu S

СтатьяПеревод в очереди

Аннотация

BACKGROUND

We evaluated the impact of a variable-pitch headless screw's angle of insertion relative to the fracture plane on fracture gap closure and reduction.

METHODS

Variable-pitch, fully threaded headless screws were inserted into polyurethane blocks of "normal" bone model density using a custom jig. Separate trials were completed with a 28-mm screw placed perpendicular and oblique/longitudinal to varying fracture planes (0°, 15°, 30°, 45°, and 60°). Fluoroscopic images were taken after each turn during screw insertion and analyzed. Initial screw push-off, residual fracture gap at optimal fracture gap reduction, and malreduction were determined in each trial. Statistical analysis was performed via a 1-way analysis of variance followed by Student t tests.

RESULTS

Malreduction was found to be significantly different between the perpendicular (1.88 mm ± 1.38) and the oblique/longitudinal (0.58 mm ± 0.23) screws. The malreduction increased for the perpendicular screw as the fracture angle increased (60° > 45°=30° > 15° > 0°). Residual fracture gap at optimal fracture gap reduction was also found to be significantly different between the perpendicular (0.97 ± 0.42) and oblique/longitudinal (1.43 ± 1.14) screws. The residual fracture gap increased for the oblique/longitudinal screw as the fracture angle increased, although the oblique/longitudinal screw with a 60° fracture angle was the only configuration significantly larger than all the other configurations. Screw push-off was not found to be significantly different between the oblique/longitudinal screw and perpendicular screw trials.

CONCLUSIONS

The perpendicular screw had a larger malreduction that increased with fracture angle, whereas the oblique/longitudinal screw had a larger residual fracture gap that increased with fracture angle.

Для цитирования: Mickley John P, Lynch Daniel J, Gordon Adam M, Roebke Austin J, Goyal Kanu S. Fracture Gap Closure and Reduction Are Affected by the Orientation of the Headless Compression Screw.. Hand (New York, N.Y.). 2023. DOI: 10.1177/15589447221150504.