Медицинская библиотека · Biomedical engineering online · 2022

A method based on non-ionizing microwave radiation for ancillary diagnosis of osteoporosis: a pilot study.

Albuquerque Gabriela, Cruz Agnaldo, Carvalho Dionísio, Mayrink Nadja, Pinheiro Bruno, Campos Antonio, Lima Josivan Gomes, Henriques Jorge, Valentim Ricardo

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Osteoporosis is a condition characterized by low bone mineral density, which typically leads to fractures and reduced quality of life. Currently, diagnostic devices used to assess this condition (e.g., dual-energy X-ray absorptiometry) are very costly, making it infeasible to meet the demand for testing in most countries. Therefore, we proposed a preclinical validation of a prototype called Osseus in an attempt to enhance osteoporosis screening tests and alleviate their costs. Osseus is a device developed to assist bone mineral density classification. It integrates a microcontroller into other peripheral devices to measure the attenuation at the middle phalanx of the middle finger, with two antennas operating at the 2.45 GHz frequency.

RESULTS

We conducted tests with plaster, poultry, and porcine bones. A comparison of the measurements of the original and mechanically altered samples demonstrated that the device can handle the complexity of the tissues within the bone structure and characterize its microarchitecture.

CONCLUSIONS

Osseus is a device that has been preliminarily validated. Ionising radiation needed for DXA tests is replaced by non-ionising microwave electromagnetic radiation. Osseus enables early detection of osteoporosis, reduces costs, and optimizes high-complexity testing referrals. There is a lack of validation studies with the reference/gold standard that are currently under development.

Для цитирования: Albuquerque Gabriela, Cruz Agnaldo, Carvalho Dionísio, Mayrink Nadja, Pinheiro Bruno, Campos Antonio, Lima Josivan Gomes, Henriques Jorge, Valentim Ricardo. A method based on non-ionizing microwave radiation for ancillary diagnosis of osteoporosis: a pilot study.. Biomedical engineering online. 2022. DOI: 10.1186/s12938-022-01038-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление