DOI 10.1111/jce.15287

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2021

Adverse events related to AtriCure EPi-Sense Coagulation Device-Analysis of the FDA MAUDE database.

Sheth Aakash R, Al Yafeai Zaki, Dominic Paari

СтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

INTRODUCTION

The AtriCure EPi-Sense Device is used for the hybrid convergent procedure, an emerging treatment for persistent atrial fibrillation (AF) and long-standing persistent AF. However, data on the AE related to the EPi-Sense device are scarce.

METHODS

Keyword "EPI-SENSE" was searched on the MAUDE database. There were 80 device reports from 2016 to 2020. After excluding reports when the device was not returned for evaluation, 79 device reports were included for final analysis.

RESULTS

The adverse events (AE) were broadly classified into 11 categories. The most common complications were pericardial effusion (25.3%), stroke (17.7%), and atrioesophageal fistula (AEF) (8.9%). Death was reported in 15 (19%) cases, 3 of which were due to pulmonary embolism, 6 due to AEF, 3 due to unknown cause, 1 due to sepsis, 2 due to events related to acute renal failure.

DISCUSSION: Pericardial effusion is a common AE reported in patients with convergence procedures and is well documented in the CONVERGE trial. The convergent procedure is unique in that the epicardial ablations are performed on the posterior wall with the radiofrequency probe directed towards the heart and away from the esophagus which in theory should reduce esophageal injuries. Despite that, a high number of AEF were noticed. Finally, there were also some reports of saline perfusion malfunction which can lead to injuries due to overheating.

CONCLUSION

This analysis of the AE related to the EPi-Sense device highlights several major AE that are previously unreported.

Для цитирования: Sheth Aakash R, Al Yafeai Zaki, Dominic Paari. Adverse events related to AtriCure EPi-Sense Coagulation Device-Analysis of the FDA MAUDE database.. Journal of cardiovascular electrophysiology. 2021. DOI: 10.1111/jce.15287.

Предметные рубрики MeSH