Anterior revision of a dislocated ProDisc prosthesis at the L4-5 level.

Jeon Sang Hyeop, Choi Won Gyu, Lee Sang-Ho

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

A case report describing our experience with anterior revision surgery for a dislocated ProDisc artificial disc at the L4-5 level, 2.5 years after the initial surgery.

OBJECTIVE

To highlight the difficulties and risks associated with the use of a repeat anterior approach for the revision of a ProDisc that has failed at the L4-5 level.

SUMMARY OF BACKGROUND DATA: As the ProDisc has a central vertical keel, more intensive vascular mobilization is required to remove the dislocated implant during surgery than that required in patients who must undergo a revision of the SB-Charité disc. To our knowledge, no prior report addresses whether the anterior removal of a ProDisc at the L4-5 level is possible in patients with severe retroperitoneal fibrosis.

METHODS

We describe our surgical experience with the anterior revision of a lumbar ProDisc that failed at the L4-5 level. Revision surgery (anterior lumbar interbody fusion with percutaneous pedicle screw fixation) was performed 2.5 years after the initial surgery because of the anterior dislocation of the artificial disc after the patient had engaged in heavy lifting. The operation was performed with a repeat retroperitoneal approach. The multiple venous injuries that occurred intraoperatively were well controlled with a 5/0 polypropylene (Prolene) suture, but the patient lost a significant amount of blood (3800 mL) during surgery.

RESULTS

The patient recovered from surgery without sequelae, and her preoperative symptoms resolved.

CONCLUSIONS

A ProDisc revision case at L4-5 level must be attempted with preparation of available venous balloon catheter for emergent bleeding control, cell saver, pulse oxymeter on great toe, and ureteral catheter. In our opinion, usage of adhesion barrier material should be considered in anterior lumbar surgery to make revision exposure easier and safer.

Для цитирования: Jeon Sang Hyeop, Choi Won Gyu, Lee Sang-Ho. Anterior revision of a dislocated ProDisc prosthesis at the L4-5 level.. Journal of spinal disorders & techniques. 2008. DOI: 10.1097/bsd.0b013e3181633a32.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.