Nitric oxide in Extracorporeal Membrane Oxygenation Sweep Gas Promotes Oxygenator Longevity: A Technical Case Report.

Stukov Yuriy, Dibert Tavenner, Vázquez-Colón Zasha, Bleiweis Mark S, Roberts Teryn R, Batchinsky Andriy, Stahl Ryan, Jacobs Jeffrey P, Peek Giles J

СтатьяПеревод в очереди

Аннотация

In the accompanying article in this issue, we reported the survival of a 4 month old infant with severe necrotizing pneumonia due to respiratory syncytial virus (RSV) and methicillin-resistant Staphylococcus aureus (MRSA), resulting in cardiogenic shock and respiratory failure, who required a total of 233 days of extracorporeal membrane oxygenator (ECMO) support. In this article, we report the function of the ECMO membrane (oxygenator). To preserve oxygenator function, nitric oxide was added to the ECMO sweep gas (sNO). Despite prolonged ECMO support, a single membrane lung functioned effectively for 5 months. Structural analysis of the polymethylpentene (PMP) oxygenator postdecannulation revealed no thrombus or fibrin deposition. This is the first known structural evaluation of a PMP oxygenator after prolonged clinical use with sNO, suggesting that sNO may prolong oxygenator longevity and maintain functional performance.

Для цитирования: Stukov Yuriy, Dibert Tavenner, Vázquez-Colón Zasha, Bleiweis Mark S, Roberts Teryn R, Batchinsky Andriy, Stahl Ryan, Jacobs Jeffrey P, Peek Giles J. Nitric oxide in Extracorporeal Membrane Oxygenation Sweep Gas Promotes Oxygenator Longevity: A Technical Case Report.. ASAIO journal (American Society for Artificial Internal Organs : 1992). 2026. DOI: 10.1097/mat.0000000000002720.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.