Peak load resistance of the JuggerKnot™ soft anchor technique compared with other common fixation techniques for large mallet finger fractures.
Аннотация
INTRODUCTION
To identify the strongest peak load resistance among four mallet finger fracture fixation methods (Kirschner wire, pull-out wire, tension-band wiring and the JuggerKnot™ (Biomet) soft anchor fixation).
METHODS
Fixation techniques were assigned among 24 specimens from six cadaveric human hands in a randomized block fashion. Peak load resistance was tested at 30°, 45° and 60° of flexion of the distal interphalangeal joint.
RESULTS
The mean peak load of tension-band wiring was 67.8 N at 60° of flexion which was most superior. The JuggerKnot™ fixation had mean peak loads of 13.35 N (30°), 22.51 N (45°) and 32.96 N (60°). No complications of implant failure or fragmentation of the dorsal fragment was noted.
CONCLUSIONS
Tension-band wiring was the strongest fixation method but was most prominent on the skin surface as seen in three specimens. The JuggerKnot™ soft anchor fixation had similar peak load resistance as k-wire fixation and pull-out wiring.